Xu Q, Jin C, Liao GX (2001) An online monitoring and alarming instrument for oil leakage based on ultrasonic emulsifying technology. Instrumentation Science &, Technology 29(5), 423-431. [In English]

Web link: http://dx.doi.org/10.1081/ci-100107234

Keywords:

Abstract: An online monitoring and alarming instrument, based on ultrasonic emulsifying technology for measuring oil leakage of circulating cooling systems is put forward in this paper. The instrument consists of a data acquisition circuit which is based on a single chip microprocessor, ultrasonic emulsifiers, photoelectric probes, and remote alarming devices. The key parts of the instrument are the ultrasonic emulsifier and the photoelectric probe. The former changes oil concentration in a sampled solution of cooling water into turbidity, and the latter detects the turbidity. Precision and reliability of the instrument depends on the function of the ultrasonic emulsifier and the photoelectric probe. Different types of oil need different input power of the ultrasonic emulsifier to yield full emulsification. Input power selections for the ultrasonic emulsifier driver, the structure of sampling tank, and the instrumentation static modeling results are presented in this paper.

 
Print    Close
 
 
   
  
 
Relevant link
Contact us
State Key Laboratory of Fire Science, University of Science and Technology of China
Jinzhai Road 96, Hefei, Anhui, P. R. China
P. O.: 230026
   
Tel:(+86)551 63601651
Fax:(+86)551 63601669
E-mail:sklfs@ustc.edu.cn
 
 
Copyright © 1990-2011 State Key Laboratory of Fire Science, University of Science and Technology of China
Tel:(+86)551 3601651 | Fax:(+86)551 3601669 | E-mail:sklfs@ustc.edu.cn | ICP: D20380176